ZSC Sensor Current Drain Check

Updated by Elizabeth Smith

This article details how to identify latent sensor issues before they result in loss of data.

Current Draw

A measurement of the current consumption of a sensor is a powerful diagnostic tool that can identify latent problems in a sensor and help maintain a sensor network.  The ZSC has an active and sleep state that has different typical currents.  Refer to the table below for typical maximum current values for sensors in the field.  As a general rule, sensors with a current draw above the maximum or an order of magnitude greater than the typical current have internal electronic problems and should be removed from the network and replaced.

ATMOS

Minimum

Typical

Maximum

Notes

ATMOS 41

Link

Awake

0.2 mA

8.0 mA

33.0 mA

Asleep

0.2 mA

0.03 mA

0.4 mA

ATMOS 22

Link

Awake

0.2 mA

8.0 mA

33 mA

Asleep

0.2 mA

0.3 mA

0.4 mA

ATMOS 14

Link

Awake

3.0 mA

3.5 mA

9.9 mA

Asleep

N/A

0.03 mA

N/A

TEROS

Minimum

Typical

Maximum

Notes

TEROS 6

Link

Awake

2.0 mA

2.5 mA

4.0 mA

Asleep

N/A

80.0 uA

N/A

TEROS 12

Link

Awake

3.0 mA

3.6 mA

16 mA

Asleep

N/A

0.03 mA

N/A

TEROS 11

Link

Awake

3.0 mA

3.6 mA

16 mA

Asleep

N/A

0.03 mA

N/A

TEROS 21

Gen1 Link

Gen 2 Link

Awake

3.0 mA

5.0 mA

16.0 mA

*Gen 2

Asleep

N/A

0.03 mA

N/A

*Gen 2

TEROS 32

Link

Awake

18 mA

25 mA

30 mA

Asleep

0.03 mA

0.05 mA

0.90 mA

HYDROS

Minimum

Typical

Maximum

Notes

HYDROS 21

Link

Awake

0.5 mA

0.5 mA

1.0 mA

Asleep

N/A

0.3 mA

N/A

LEGACY

Minimum

Typical

Maximum

Notes

5TE

Link

Awake

0.5 mA

3.0 mA

10 mA

Asleep

N/A

0.03 mA

N/A

5TM

Link

Awake

0.5 mA

3 mA

10 mA

Asleep

N/A

0.03 mA

N/A

GS3

Link

Awake

0.5 mA

3.0 mA

30 mA

Asleep

N/A

0.03 mA

N/A

10Hs

Awake

0.5 mA

12 mA

20 mA

Link

Asleep

N/A

0.03 mA

N/A


How did we do?


Powered by HelpDocs (opens in a new tab)